This article addresses many of the more common error messages and issue symptoms seen on the OBR 4600 and OVA 5100 series systems, each as its own section with potential resolutions and/or diagnoses. Common causes of many of these errors are from damaged or dirty ports / connectors / calibration artifacts, or from software mismatches or settings. This article prioritizes self-resolvable errors and warnings. However, other errors stem from hardware degradation for older instruments that will often need a Luna Support agent (see below) for additional assistance. As such, if the steps here don't help, please document all resolution steps taken and proceed to contact Luna Support. Resolution can include scheduling a remote session to further diagnose the instrument and / or proceeding with an RMA for repair.
Basic Troubleshooting:
Before working through a specific error below, attempt the following to potentially rule out a simple cause of the problem, and to gather additional data on the path to failure:
- Power cycle the instrument, and then the software as well. While doing so, confirm the instrument's LEDs light normally, as follows:
- Upon startup, all four LEDs will flash on briefly, and then only the amber Power LED will remain on.
- Upon software startup, the green Source LED will also illuminate.
- If no LEDs illuminate, confirm the power cable is connected. Note whether the rear cooling fan is spinning as an additional confirmation of power to the instrument.
- When taking a scan with the OBR, the blue Measure light will illuminate briefly and turn off, followed by the white Transfer light illuminating during data processing. With the OVA, the blue Measure light will illuminate briefly and turn off.
- Scan range affects scan time. If the Measure light stays on for more than a second (at max range), this could indicate a laser control error that will likely result in "Timed out..." errors (see section below).
- If experiencing connection issues, try a different USB cable and a different USB port on the controller PC.
- If you have another controller PC available, try connecting from it.
- Note your software version (shown in the About window). You'll need it if you contact Support.
USB connection lost, or the OBR is not detected; "A communication error has occurred in the system."
What it usually means: The USB link between the instrument and the controller PC dropped or never came up. The OBR software does not reconnect on its own after the cable is disconnected.
Try this:
- Reconnect the USB cable, then close and restart the OBR / OVA software.
- Try a different USB cable, and check the USB connector on the rear of the instrument for damage or a loose fit.
- Confirm your software version matches your instrument. Older units can use an earlier release (for example, v3.13.0, while 2021 and newer OBR units require v3.16.2). The first two digits of the serial number corresponds to the last two digits of the year of manufacture. If you're unsure, contact Luna Support with the serial number. The latest software should work with older instruments but compatibility is not guaranteed - no known issues have been reported with either the OBR or OVA software.
- Open Windows Device Manager. If the instrument appears as an unidentified device, reinstall the driver by following Windows USB drivers for OBR 4600 or OVA 5000. This is common with replacement controllers.
"The OBR (or OVA) software is not enabled in the attached instrument."
What it usually means: While this could be a problem with the system's on-board memory, it may also be related to a software mismatch - see the preceding section to check compatibility and troubleshoot.
Contact Support if the software is up-to-date. A remote session can check the on-board memory and basic laser communications, as well as start an RMA if hardware is suspected to be the cause.
The USB driver is blocked after a Windows update
What it usually means: A recent Windows update is enforcing driver rules that block the instrument's USB driver. Luna is investigating a permanent fix.
Try this:
- Make sure up to date drivers are installed. Step by step instructions here: Windows USB drivers for OBR 4600 or OVA 5000.
- See if you are affected by the Windows Driver Enforcement Policy and apply the Temporary Fixes detailed here: Windows USB Driver Enforcement (OBR and OVA)
- If these methods don't work, try installing the software on another PC, possibly running a different version of Windows.
"Timed out waiting to acquire data"
What it usually means: The instrument didn't return measurement data in time. The laser was instructed to sweep but data did not reach the processor. Occasional timeouts can be resolved by letting the instrument warm-up sufficiently, re-starting the software, and attempting another measurement sweep. Frequent, repeated timeouts, especially on older instruments, often point to wear in the laser module (control board, opto-block or power supply), or, less commonly, a non-functioning component in the acquisition and processing circuitry.
Try this:
- Power-cycle the OBR, then restart the software.
- Note how often the error occurs, which operation was being attempted (alignment, calibration, a scan) and the details of the calibration / scan parameters.
- Also make note of any recent changes, such as relocation, shipping or a new controller PC.
- If the blue Source LED light stays on for longer than a second prior to seeing this error, this is an indication of a laser control problem. It can be double-checked with a remote session but likely needs an RMA to fully remedy. Power cycling will often temporarily fix the problem.
Contact Support if the timeouts continue. A remote session can check laser health and raw signals to determine whether the laser module needs repair or replacement, or if there is an optical problem with one or more of the signal channels coming through to the photodetectors, or in the processing circuitry once the signal is acquired.
Communication errors that persist after the USB checks
What it usually means: If the USB cable and driver all check out, and especially if the instrument was previously working, the laser in an unexpected state can trigger calibration errors that look like communication errors. This usually needs to be escalated to a Luna Support ticket and may turn into an RMA.
Try this: Contact Support with the exact error text and the steps you've already tried. Support may recommend returning the unit for evaluation.
"Laser Not Ready" warning text
What it usually means: The laser needs 5-10 minutes in a powered on state to reach and keep a stable temperature. If it is taking an unusually long time to warm up and the warning to disappear, or if it suddenly turns on after the instrument has been on for a longer period of time, that usually indicates a problem with the Thermoelectric Cooler (TEC), or the overall temperature control loop.
Try this:
- Check to see whether the fan is spinning properly in the rear of the instrument chassis. On rare occasions, a gray communication cable has been known to slip out of place and contact the fan blades. This does not damage the blades but can interrupt fan operation.
- If it is not spinning, turn off the instrument and check for the cable using a flashlight. If confirmed, use a thin tool (such as a screwdriver) to gently re-position the cable away from the fan.
- Ensure proper room behind the instrument for fan exhaust, especially in warm environments.
Contact Support with the details on when the text appears (on start-up, or after a long period of use), as well as if the instrument is being operated in unusual environmental conditions. A remote session can verify the temperature of the laser and a support agent can then begin an RMA if necessary.
"Wavelength cal failed" warning text
What it usually means: The laser sweep has drifted in wavelength, and software corrections can not be performed. A common sign is that calibration still succeeds over a narrow or particular range (for example, 20 nm near 1535 nm) but fails over the full sweep.
Try this:
- Clean the front-panel port (see the cleaning procedure in your OBR User Manual).
- Inspect the gold reflector and replace it if it's damaged or dirty. You may be able to temporarily use a flat-cleaved fiber end in a pinch (see discussion here).
- Re-align.
- Attempt calibrations and subsequent scans in sets of ranges, such as 7, 10, 15, or 20 nm on the low, middle, or high ends of the overall range, and note if the error is consistently seen in one range but not others. Record the findings.
Contact Support if calibration still fails. A remote session can confirm wavelength drift. A Luna Support agent can discuss repair-ability of the laser under an RMA.
"Out of Alignment" after calibration, or calibration getting harder over time
What it usually means: The calibration routine attempts to balance the polarized signals detected on the reference path but can have difficulty of there is some unexpected problem in one or both of the signals. It's possible for a poor connection or problem with the gold reflector (in reflection) to be the root cause. If calibration has become gradually harder over months, the laser, particularly the opto-block, may also be failing.
Try this:
- Replace the gold reflector, if possible. If not, attempt to calibrate with a flat-cleaved jumper on the termination end. The reflector does not need to be at or near 0 dB (see the first two paragraphs here). Success with this method would indicate a need to replace the reflector.
- Clean the front-panel port (See here and the OBR User Manual, page 195).
- Re-align and re-calibrate.
Contact Support if the error persists. A remote session with a Luna Support Agent can check laser operation and the polarized raw signals for other possible causes.
OVA: "exceeded PDL tolerance during external calibration"
What it usually means: The problem is usually in the external calibration path, not the instrument.
Try this:
- Fully uncoil the calibration paddles and let them settle, then retry.
- Try a different set of paddles, if available.
- Clean, re-seat and check the Source and Detector connections if able.
OVA: "high loss" during reflection calibration
What it usually means: An external component, such as the bulkhead adapter, calibration reflector or calibration DUT, is dirty, damaged or not fully connected.
Try this:
- Disconnect, inspect and clean each ferrule before reconnecting the calibration reflector.
- Repeat the calibration.
- If needed, try a different calibration reflector, if available.
- If you have a license for the OFDR Software, a scan with nothing attached to the Source port can point to where the problem may lie, such as showing a very bright (<-53 dB with a 0.5m integration width) Source port reflection peak.
"Detector Saturated – Reduce Gain"
What it usually means: Too much light is reaching the detector for the current gain settings, often from a highly reflective DUT.
Try this:
- Confirm the gain is set to 0 dB. Use the highest Gain value that does not result in the above warning.
- Check the reflection levels of your DUT and connections, particularly if the DUT reflections are focused on a narrow wavelength band.
Contact Support if the message persists at 0 dB gain. There may be a parameter setting that can be adjusted to resolve the error.
OVA 5000: laser failing ("Error 13")
What it usually means: The OVA 5000 has reached end of life and is no longer supported, so its laser can't be repaired. A Support Agent may be able to perform a brief remote session to verify the laser will not communicate in a basic way.
What to do: Contact Support for verification. Be prepared to contact your Luna sales representative about a replacement instrument.
Controller laptop won't boot
What it usually means: While Luna provides controller laptops with our instruments, the care and maintenance and troubleshooting of these controllers is primarily up to the end user. For the OBR 4600 and OVA 5100 product lines, most modern computers running Windows will exceed the original system specifications and can be set up to run the software needed. For the OBR and OVA software, you may need to submit a Support Ticket to get the latest version. You will also likely need to update drivers after the software is installed (see The USB driver is blocked after a Windows update section above).
Try this: For systems under warranty, contact Luna Support directly by submitting a ticket. For systems out of warranty, you may contact your Luna Sales Rep about ordering a replacement laptop, contact the laptop manufacturer directly for support and service, or replace the laptop through other means.
Still need help?
Submit a ticket and include your product model, serial number, software version, the exact error text and the steps you've already tried. For more on what to include, see Getting Started with Luna Customer Support. The next two sections are brief summaries of remote sessions and what to do when there is a likely hardware issue.
Preparing for a remote session
If the steps above don't resolve the error, Support will usually schedule a remote session. Support no longer uses the older self-diagnostics program; instead, a System Diagnostic program is used to check laser functioning and raw signal values. To prepare:
- Make sure TeamViewer is installed on the OBR controller laptop and that the laptop can reach the internet over wired Ethernet or strong Wi-Fi. If your controller laptop must stay offline, discuss with Support to make other arrangements. Remote Support may be limited in these cases.
- Have the instrument powered on and connected to the controller laptop.
- Before the session, make sure relevant details of the setup and observed errors are shared via the Support ticket.
- Plan to be available during the session. Your agent may let you know that they can check the issue on their own, without someone present locally, depending on the issues. TeamViewer has an audio feature, but a Microsoft Teams meeting can also be arranged to handle an audio call.
For full requirements, see the Remote Troubleshooting Guide.
When it's likely a hardware issue
Some errors point to a fault that needs repair at Luna rather than a fix you can make on site:
- Repeated laser-related errors, such as frequent timeouts or wavelength calibration failures that persist after cleaning and re-calibrating.
- Communication errors that remain after the cable, port and driver checks.
- Signal checks that show a light path has gone dark, indicating a fiber break within the system.
In these cases Support may suggest starting an RMA. Repairs are subject to a minimum repair fee unless covered by warranty; Support will confirm the cost before work begins. For warranty terms, see Warranty Coverage for Luna Innovations Products.