Polarization Extinction Ratio (PER) describes how much light crosses from the intended polarization axis of a device into the unintended one. In the OVA time domain response, this crossover appears as small peaks between the two large peaks that correspond to the fast and slow transmission axes. Crossover can result from misalignment of the PM fiber axes to the TE/TM axes of a waveguide, or from a defect in the waveguide or fiber.
The procedure below has been demonstrated by Luna on the OVA and uses careful placement of the time domain filter on two copies of the same scan. This procedure applies to the OVA 5100 and OVA 5113 models, as well as the legacy OVA 5000 and 5013 models. It is performed in the OVA Control software rather than the Polarization Analysis Software, because it relies on storing the same data set in multiple traces.
Measurement procedure
Step 1: Scan the DUT
Find the DUT length, take a scan, and set the time domain filter window. Then display the Time Domain Amplitude Response. You should see two large peaks for the fast and slow transmission axes, plus a couple of minor crossover peaks between them.
Step 2: Copy the trace
In the OVA Control software, copy trace A to trace B so that both traces hold the same data set.
Step 3: Set the time domain windows
On trace A, set the time domain window narrow so that it passes only the two small inner peaks.
On trace B, set the time domain window wide so that it passes light from all four peaks.
Step 4: Read PER from the Insertion Loss
Display Insertion Loss for both traces. You should see a large difference between the two curves. The gap between them is the ratio of light passing through all states of the device to the light that crossed over, which is the PER. This is a very good approximation when the ratio is large.
Use the horizontal cursors to measure the difference between the two curves. In the example below, the difference between the two curves was measured at PER = 17.8 dB.